Basic Ultrasound Physics for EM 

Jason T Nomura
Подписаться 9 тыс.
Просмотров 311 тыс.
50% 1

CORRECTION: 0:29 Megahertz = million hertz so 2 Megahertz is 2,000,000 hertz.
CORRECTION: 2:26 Speed of sound though soft tissues ranges from 1450 m/s (adipose) to 1580 m/s (muscle) and most ultrasound systems assume a default speed of sound of 1540 m/s for "tissue".
The video has an error of 1080 m/s listed (totally missed it during editing, thank you Ken Rolt for catching it).
Video reviewing basic ultrasound physics, ultrasound modes, artifacts, probes types, and imaging planes with orientations.
website: takeokun.com
youtube: / itakeokun
twitter: @takeokun
For informational and educational purposes only. Does not replace formalized medical education and/or credentialing. Presented "as is".

Опубликовано:

 

27 июн 2024

Поделиться:

Ссылка:

Скачать:

Готовим ссылку...

Добавить в:

Мой плейлист
Посмотреть позже
Комментарии : 107   
@irahi9501
@irahi9501 3 года назад
On RU-vid settings slow the video speed to 75, it works perfect
@sidology1.0
@sidology1.0 3 года назад
exactly what i was thinking! Lol a lot of information to absorb at what seems like lightspeed
@Ravi_Qant
@Ravi_Qant 11 месяцев назад
Meanwhile me, listening to it at 1.25x speed. ☠️ Edit. This is making no sense without animation.
@ellasnow6239
@ellasnow6239 7 лет назад
Thank you!! That condensed hours worth of pre-reading for my POCUS course!! You're a legend.
@gregorycostello239
@gregorycostello239 2 года назад
I am a physician ready to get started in point of care ultrasonography,. What a great review of the fundamentals! This really made my studying make much more sense. Your diagrams and animation are incredibly informative Thank you!
@filipamk1011
@filipamk1011 3 года назад
Thank you for taking the time to do this video! Extremely helpful 🙂
@MartaCrowe
@MartaCrowe Год назад
Thank you for making this! I've been designing in ultrasound for 2 years and this is the most helpful video on US physics I've seen yet.
@facesunderwater
@facesunderwater Год назад
This video is wonderful. You explain attenuation of high & low frequencies very nicely. Thanks!
@khudhurm7709
@khudhurm7709 7 лет назад
This video has answered most of my questions on ultra sound imaging (basics), Thank you :)
@estherssebbowa3043
@estherssebbowa3043 3 года назад
Great video - a summary of almost everything in under 20!
@thecerebrum.
@thecerebrum. 3 года назад
This is very well put. Thank you
@rashmiks1564
@rashmiks1564 3 года назад
Thank you very much sir, very informative for understanding the basics of ultrasound 🙏🏻
@user-qe5uw9cj5f
@user-qe5uw9cj5f 4 года назад
Super helpful video! And I think the speed was perfect
@Sand24
@Sand24 7 лет назад
Good job explaining...i am looking into this career and learning a lot from these videos.
@hannahirvin8567
@hannahirvin8567 2 года назад
I have done ultrasound a few years and I had gotten foggy on the physics. This covers the most important physics as fast as possible.
@tammieannie6135
@tammieannie6135 7 лет назад
Great video! Thank you very much
@vissompelotas
@vissompelotas 3 года назад
Perfect and fast revision!!!
@tomodude6542
@tomodude6542 4 года назад
This video is excellent. Thanks very much.
@erayk96
@erayk96 3 месяца назад
Perfect video. Completely answered all my quriosity about how it works. Thank you.
@motivasivideoinspirasi8953
@motivasivideoinspirasi8953 3 года назад
Thanks for the great lecture.
@Nitrox-.
@Nitrox-. 5 месяцев назад
This is an amazing video. Just quick fire, well explained most relevant information. Helped me so much to compensate for the horrible lecture I got. Somehow my professor spent 4 hours explaining mostly the same things and I didn't understand a thing.
@avieleliyahu6199
@avieleliyahu6199 5 лет назад
Thanks! really clear and helpful :)
@maanceto2
@maanceto2 7 лет назад
This video is incredibly useful!
@TakeoKun
@TakeoKun 7 лет назад
Thank you, hope it continues to help. feel free to share with anyone you think might benefit or be interested.
@lolwaha943
@lolwaha943 2 года назад
You literally saved my future!! Thanks alot❤️❤️❤️
@fadeskywards1245
@fadeskywards1245 4 года назад
Very helpful. Thank you!
@papaschlumpf1390
@papaschlumpf1390 2 года назад
My man is speed running a whole topic of medical diagnosis in a sub 20 run
@fjs1111
@fjs1111 4 месяца назад
Jason, that is very interesting. I've worked with Phased Array Radar elements and RF, this is basically phased array for ultrasound. Amazing!!
@fcrick6967
@fcrick6967 2 года назад
Superb - thank you!
@fadyalhosain9170
@fadyalhosain9170 4 года назад
very good job,thanks alot sir
@jeremygerber2996
@jeremygerber2996 3 года назад
Thanks, great video!
@bundled-micro-coaxial-cable
Nice video, thank you !
@d-293-youssefehab9
@d-293-youssefehab9 Год назад
Amazing explanation
@javedahmad2234
@javedahmad2234 5 лет назад
This is an excellent video. Thank you so much. I am using it for teaching purpose. It saved a lot of my time. Very nice presentation, very well prepared and very well presented, I would suggest to be a bit slower so that the trainees grasp it without too much pause. Thanks a lot for the efforts.
@redietnigussie8383
@redietnigussie8383 Год назад
Thank you, it helps a lot
@trayanc1496
@trayanc1496 3 года назад
this was so beneficial
@glob3242
@glob3242 3 года назад
Well done!
@ravindren2915
@ravindren2915 5 лет назад
Thank you so much
@user-ih4yh9ww2u
@user-ih4yh9ww2u 3 года назад
Thanks so much!
@kenrolt8072
@kenrolt8072 7 лет назад
Speed of sound in tissue, at 2:24 in the video, is not around 1040 m/s. It's in the range of 1450 to 1580 m/s depending on the tissue type. Ultrasound machine default speed assumed for soft tissue is 1540 m/s because it's a weighted average of what's most likely to be encountered in a scanning session.
@TakeoKun
@TakeoKun 7 лет назад
Thank you, you are right, not sure how that made it through editing. thank you for catching it; I have added a note to the comments section for the correction.
@kenrolt8072
@kenrolt8072 7 лет назад
Thanks for replying and for the videos.
@slippyg
@slippyg Год назад
Yeah I noticed that too.
@Applxy354
@Applxy354 3 года назад
I am less worried about speed of sound and more about the speed of your video.
@lulusariwu6413
@lulusariwu6413 Год назад
Amazing!
@ronidaffan5904
@ronidaffan5904 2 года назад
Thank you so much !!!!
@radiofollowAlotbi
@radiofollowAlotbi 4 года назад
very good sir
@kentkeatha6728
@kentkeatha6728 5 лет назад
Thanks sit
@badbaldboujee7949
@badbaldboujee7949 7 лет назад
Is there a transcript for this video?
@sravanipeddireddy1967
@sravanipeddireddy1967 4 года назад
Little bit explain slowly means it's better to more understanding....
@bashirmalik6627
@bashirmalik6627 3 года назад
Machines are calibrated at the assumption of the speed of sound in soft tissues at 1540 m/sec instead of at 1040 which you said at 2.35 minutes at your lecture, please review. Thank you
@TageldinAbogrounDr
@TageldinAbogrounDr Год назад
Thank you very much for your helpful and formative video ❤🎉 but an average value of speed of sound in soft tissue is 1540 m/s 😊
@jenngzz11
@jenngzz11 7 лет назад
Interesting. 😍
@zaeemtahir3020
@zaeemtahir3020 7 лет назад
is ultasound has low amplitude?
@robogamist8250
@robogamist8250 6 лет назад
Hi, very informative video. What font have you used for your text? Many thanks.
@TakeoKun
@TakeoKun 6 лет назад
Thank you. I used Helvetica light which was the default for the version of keynote used to create the slides.
@marianelkassab3066
@marianelkassab3066 5 лет назад
Thx
@ahmadalbassal4179
@ahmadalbassal4179 7 лет назад
thank alot
@denisela3403
@denisela3403 5 лет назад
That's like over half of Edelmans book in 17 min🤯😱
@irahi9501
@irahi9501 3 года назад
Too fast
@kenrolt8072
@kenrolt8072 7 лет назад
Also be careful with BART for Doppler colors. The speed colorbar shown on US displays will tell you unambiguously if the flow is towards or away from the transducer by looking at the color AND the corresponding positive or negative sign for the speed value. Don't just assume BART gives the proper interpretation.
@crikeymos22
@crikeymos22 6 лет назад
Agreed. These types of mistakes in teaching are unforgivable as it confuses the trainee.
@murtazahassan6806
@murtazahassan6806 5 лет назад
Ken Rolt new addition to my knowledge.thanks
@delux415
@delux415 7 лет назад
There is a error! 2MHz is 2 000 000 Hz, not 200 000Hz...
@Objectra
@Objectra 5 лет назад
That caught my attention as well.
@AKASIFAK
@AKASIFAK 9 месяцев назад
Just to point out 200,000hz isn't equal to 2 mgz as you said in the beginning of the video. The correct conversion is... 2000000hz is actually 2mhz. Apologies if I misunderstood something , but great video nonetheless!
@bryanbinoy9732
@bryanbinoy9732 4 года назад
TKS BRIEF EXPLANATION
@pinklemonade3919
@pinklemonade3919 4 года назад
Soft tissue is typically assumed to be 1540 m/s, not 1040 m/s. Good video though!
@Hendreh1
@Hendreh1 2 года назад
2MHz is 2.000.000Hz. Very basic stuff.
@derrickniu6220
@derrickniu6220 2 года назад
2MHz=200,000Hz? or 2,000,000Hz
@SergeantTubesocks
@SergeantTubesocks 4 года назад
We use A scan
@murtazahassan6806
@murtazahassan6806 5 лет назад
Awesome man...loved the lecture.only faced one problem...you talk so fast...I had to reduce the speed...but the lecture was awesome...will look the rest of the lectures now to learn what many teachers usually never talk about.... In indopak region, the teachers never teach you just so that no one can replace them.they will rob you off your money but you will get nothing at the end of 6 months unfortunately.... Once I mentioned the B mode and A mode modalities and the teacher got furious and he said there is no such thing like B mode or A mode. Now what to do with such a person. 90% of the echo reports formed at even the major hospitals in kpk, Pakistan are damn damn inaccurate.quackery is practiced at its worst but I refuse to practice any such quackery.... Let me amaze you now 😵 it takes atleast 20-30 min to perform one examination but here in my country it takes 2-3 mins only.... Now decide what kind of echo report would that be...thanks to RU-vid that gave me teachers from around the world
@aasiamalik1220
@aasiamalik1220 3 года назад
I m too from Pakistan, Karachi . I take more than 20 minutes to complete one normal scan of heart but if there are abnormal findings it take more than 30 minutes and sometimes more than an hour. I work in tabba heart institute where we follow almost all international protocols to make sure the study should be perfect.
@anasuyamukherjee7872
@anasuyamukherjee7872 3 года назад
Same situation in India too. U mention anything that your professor didn't mention while he/she was explaining something ,they take it all in their ego and humiliate you with unrelated things to make sure you never speak up again. Even though you literally have to beg, for them to take any form of class and even if they ever teach anything that comes down 'how to break a student's morale'. But, this is not the case for very senior professors though.
@kratex100
@kratex100 3 года назад
Can anyone tell me if it is possible to stop ultrasound waves that are being directed at me by my upstairs neighbors?
@michaelziyizeng2216
@michaelziyizeng2216 3 года назад
2Mhz is not 200,000 is 2 ,000, 000hz , please correct it.
@benclark3513
@benclark3513 4 года назад
I love this I like ultrasound the ultrasound is so legit :)
@johnterry6541
@johnterry6541 4 года назад
Why would you mess up such a good video by talking so fast. What's the hurry! I'm looking for a 0.5x speed.
@fadiaezzeddine7271
@fadiaezzeddine7271 5 лет назад
Very fast but good information
@katvadana6728
@katvadana6728 5 лет назад
way too fast he needs to slow down so we understand better take a breath
@mariem6516
@mariem6516 3 года назад
Oh my goodness it would help if you would talk a little slower. What’s the rush? Other than that it’s great info!
@mymoonly
@mymoonly 4 года назад
I thought the average speed in soft tissue is 1.540 m/s
@OlenKukat
@OlenKukat 4 года назад
Mondblume Mystic Yeah also MHz is million... He said 2MHz was 200,000 :/
@whitepigine7672
@whitepigine7672 4 года назад
OMG
@sandracade641
@sandracade641 2 года назад
Okay 4
@meocoi9506
@meocoi9506 5 лет назад
Very useful information but talk too fast and no bread in between.
@mansiudayjoshi
@mansiudayjoshi 3 года назад
Playback speed infinity Thanks though
@sandracade641
@sandracade641 2 года назад
OKC ✅☺️ a
@mikemartinez9041
@mikemartinez9041 6 лет назад
I’m pretty sure through tissue is 1540 and no 1040.
@ramasitadevikunadharaju3434
@ramasitadevikunadharaju3434 6 лет назад
Mike Martinez 5
@dv9096
@dv9096 5 лет назад
you're right! it's always been 1540m/sec for soft tissues
@najeebullah587
@najeebullah587 5 лет назад
The speed of sound in tissue is 1540m/s but,it changes with the change in the density of tissue . The higher the density the faster will be the wave and vice versa
@javedahmad2234
@javedahmad2234 5 лет назад
@@najeebullah587 agree and thanks for the useful comment
@kumarn4674
@kumarn4674 5 лет назад
Sound in tissue on average 1540 m/s or 1.54 mm/us
@TakeoKun
@TakeoKun 5 лет назад
That was an error when the recording was done, there is a notation in the description with the correction.
@kumarn4674
@kumarn4674 5 лет назад
@@TakeoKun Ok great, overall good review!
@abcdefg5185
@abcdefg5185 4 года назад
talks so fast. whats the hurry ??????
@tacoram3834
@tacoram3834 4 года назад
Good job explaining, but wish you didn't talk so fast
@justineanyigba9831
@justineanyigba9831 Год назад
Can u please just slow down
@MuhammadArif-gg4zu
@MuhammadArif-gg4zu 8 месяцев назад
Un necessary Very fast
@anukvangelderen2457
@anukvangelderen2457 3 года назад
you talk so fast.... loved the fact that the video explained everything but wasn't 2 hours long, but please talk slower next time.....
@pastblaster3285
@pastblaster3285 3 года назад
Go to the wagon wheel / settings / lower right / and look for playback speed and change it to how you want it to be ..............
@josephokonkwo6237
@josephokonkwo6237 Год назад
TOO FAST
@daisyroots8926
@daisyroots8926 3 года назад
Please speak more slowly 👍🏻
@Jasmineliketheflower
@Jasmineliketheflower 5 лет назад
I am sorry but this was hard to understand only due to your speech... your speaking style in this video is not clear at all. If this was written text there would be no punctuation. I had to slow it down and rewind quite a few times too.
Далее
Ultrasound Physics and Instrumentation
48:06
Просмотров 122 тыс.
Ultrasound Basics
36:39
Просмотров 312 тыс.
Задержали в аэропорту
00:56
Просмотров 164 тыс.
Bedside Ultrasound Basic Cardiac US
19:22
Просмотров 608 тыс.
Ultrasound Transducer Manipulation
7:21
Просмотров 430 тыс.
EEVblog #1315 - Ultrasound Probe Extreme Teardown!
18:32
Introduction EFAST Focused Ultrasound Examination
27:13
How to see with sound - Jacques S. Abramowicz
5:16
Просмотров 366 тыс.
Ultrasound Physics - Explaining Doppler
3:51
Просмотров 261 тыс.
MRI: Basic Physics & a Brief History
25:52
Просмотров 801 тыс.
Ultrasound || Basics and Beyond || Dr. Abhishek Jha
20:24
Level 1 - Ultrasound Physics
31:12
Просмотров 14 тыс.
Zonare ZS3 Basic Knobology
15:02
Просмотров 24 тыс.