Тёмный

BW Technologies GasAlert Max XT II Gas Calibration Guide by Ideal Calibrations 

Ideal Calibrations
Подписаться 2,3 тыс.
Просмотров 163 тыс.
50% 1

Опубликовано:

 

10 сен 2024

Поделиться:

Ссылка:

Скачать:

Готовим ссылку...

Добавить в:

Мой плейлист
Посмотреть позже
Комментарии : 59   
@niloyu105
@niloyu105 3 года назад
Thanks for this very helpful video watching here Al Khafji Saudi Arabia from Philippines...
@Dianapadron81
@Dianapadron81 2 года назад
You have a very pleasant voice. You should make more instructional videos.
@fuckpoliticalcorrectness9849
@fuckpoliticalcorrectness9849 9 месяцев назад
Thank You For the helpful video straight to the point
@Idealcalibrations
@Idealcalibrations 9 месяцев назад
Thanks for the kudos, let me know if you run into anything you need help with!
@MOHAMMADJAVED-jb4pk
@MOHAMMADJAVED-jb4pk 4 года назад
Thanks Jeams, i appreciate your reply, yes i found at the inlet pump's case edge little bit pressed to inside let me fix it than i will get back to you
@clownbaby8209
@clownbaby8209 7 лет назад
thank you for your simple & easy to follow tutorial
@melquiadesalmeida1133
@melquiadesalmeida1133 2 года назад
Muito bom esse vídeo, muito prático.
@kaoilfieldchannel838
@kaoilfieldchannel838 4 года назад
Thanks for video helpful for us
@jamesnelson7372
@jamesnelson7372 3 года назад
Hello sir i try to follow your procedure but whe i enter in calibration mode i get notification shown ( IR LOCK ON) HOW TO UNLOCK?
@Idealcalibrations
@Idealcalibrations 3 года назад
Hey James, you need to hook it up using the IR link from BW and the BW Fleet Manager software. Someone must have enabled that in your config.
@inyomanaryasuganda4397
@inyomanaryasuganda4397 2 года назад
thank you... it is nice
@lobutayasgeneration825
@lobutayasgeneration825 2 года назад
How to calibrate if Cal. lock or IR Lock appears on the display, if available calibrate by manual
@Tonyskiable
@Tonyskiable 8 лет назад
Hi My Gasalert Max xt ii is not blocking during start up other pump function works well and I have checked on the software the blocking function is activated. What should I do
@Sophia_Mariano
@Sophia_Mariano 7 лет назад
hi, just want to ask why i can not calibrate my gas monitor even i do all your procedure. it always show h2s error and co but it reaches the span value. always the 2 sensor showing error thats why it indicates a failed calibration. tnx in advance
@Idealcalibrations
@Idealcalibrations 11 лет назад
Hey Michael, no this one does not. The unit you're thinking of is the BW Technologies GasAlert Quattro monitor. It's a diffusion unit. Go to the IdealCalibrations page and check out the Quattro page for some more info. RU-vid won't let me link directly.
@BhaskarRoy-f1p
@BhaskarRoy-f1p 14 дней назад
Hi James, during calibration, upon connecting the span gas cylinder, it's saying "Pump Off" and calibration is failing. Can you advise.
@Idealcalibrations
@Idealcalibrations 14 дней назад
@@BhaskarRoy-f1p hey there, sounds like you're using a fixed flow regulator, does it have an on/off knob? If so, make sure the valve is open the entire time, don't hook it till the unit till it is open. If I were you I'd recommend using a demand flow regulator instead
@BhaskarRoy-f1p
@BhaskarRoy-f1p 14 дней назад
@@Idealcalibrations Yes, on/off knob with a regulator. Do I have keep it fully open? I just left it like partially open. Kindly advise.
@Idealcalibrations
@Idealcalibrations 14 дней назад
@@BhaskarRoy-f1p fully open, it's not a flow adjust valve, you want it either all on it all off
@NAPSTEROWN
@NAPSTEROWN 2 года назад
I ve dropped mine it the manhole in the water and it doesnt want to open/activate! What can i do?
@Idealcalibrations
@Idealcalibrations 2 года назад
I'd send it in to a shop at that point. Most likely you've got some water damage on the board. Welcome to send it to us if you'd like. Number for the shop is 734-956-0539
@medaminqhse6343
@medaminqhse6343 3 года назад
Thanks for the explanation and I look forward to working with... 😊😉👍🏼⚒️☺️😃💬👀💯👌
@Idealcalibrations
@Idealcalibrations 3 года назад
Thanks for the kudos! Let me know if you ever have any questions I can help with
@michaelb4041
@michaelb4041 11 лет назад
Does this have the green led on the top to show that it has been bumped and cal?
@tranletrung3045
@tranletrung3045 Год назад
How much gas used for 1 calibration please?
@Idealcalibrations
@Idealcalibrations Год назад
The calibration takes about 2 mins and the flow rate is close to 0.5lpm, so I generally assume 1 liter of gas per calibration as a loose estimate.
@MOHAMMADJAVED-jb4pk
@MOHAMMADJAVED-jb4pk 4 года назад
HI, Jeams my machine is not Geting PUMP OK, what to do?
@Idealcalibrations
@Idealcalibrations 4 года назад
Hey Mohammad, sounds like your pump block is failing. Check the seal where your input and filter holder meet the back case. Usually if you unscrew this and remove it, you'll seem some damage to the back case. 95% of the time that's the problem and you either need to fix or replace that back case. Let me know if that doesn't fix it
@jenniferhansen9867
@jenniferhansen9867 2 года назад
Needed
@fenotriadi5255
@fenotriadi5255 4 года назад
I would like to ask if this method call as calibration? Because my distributor said that I need micro dock to calibrate. Your method, they call only a bump test. Please advice.. Thank you..
@Idealcalibrations
@Idealcalibrations 4 года назад
Your distributor is incorrect. You can both bump test and calibrate a detector manually, with just the gas and the regulator. All a calibration requires is the steps in the video. Sounds like they're trying to sell you a dock.
@fenotriadi5255
@fenotriadi5255 4 года назад
@@Idealcalibrations tq for the answer..
@argenttillett4635
@argenttillett4635 8 лет назад
Am trying to calibrate my meter but am using a 0.2 flow meter. Would that give me a false reading
@Idealcalibrations
@Idealcalibrations 8 лет назад
Unfortunately that does sound too slow. If it's a pumped unit, it's going to do less than nice things to the pump motor over time as well, because it's not flowing enough gas for the pump.
@MrNoenevarez
@MrNoenevarez 2 года назад
Where can I purchase the regulator and calibration gas tank?
@Idealcalibrations
@Idealcalibrations 2 года назад
Hello, you can pick the gas and regulator up here: idealcalibrations.com/calibration-gas-and-accessories/?_bc_fsnf=1&Model=GasAlertMax+XT+II
@muhammadazhar2131
@muhammadazhar2131 9 месяцев назад
How about IR LOCK ON?
@Idealcalibrations
@Idealcalibrations 9 месяцев назад
Sounds like it's set for IR lock for calibration. Hook the unit up to Fleet Manager using the data link and remove that check box in the config and you should be set.
@zedgerton73
@zedgerton73 3 года назад
For the one person that reads the comments you might find this funny. I was a green laborer running sanitary sewer. I had to calibrate this exact machine to sniff out the manhole. I thought that bottled gas was oxygen so my stupid ass took a few huge wiffs of it…I was like damn this oxygen smells like shit, I had 3 people smell it before I finally read the contents 💀
@Idealcalibrations
@Idealcalibrations 3 года назад
Dude no 😂 Yeah definitely not the sort of thing you want to be breathing. At least the smell lets you know the Hydrogen Sulfide is fresh
@katonbagasf9753
@katonbagasf9753 8 лет назад
Hi i wanna Ask , what the function from block inlet ?
@mrxalos
@mrxalos 4 года назад
To test that Puma is working
@marcoruiz3345
@marcoruiz3345 10 лет назад
How can I set de custom start-up message? Thanks.
@Idealcalibrations
@Idealcalibrations 10 лет назад
Hey Marco, you need to go into the options using the Fleet Track software from BW Technologies as well as the USB cord you buy from them. They're about $75 from what I remember off the top of my head. Once you're in there it's fairly easy. Email me if you have any questions from there.
@jenniferhansen9867
@jenniferhansen9867 2 года назад
Do you have a link to the parts
@Idealcalibrations
@Idealcalibrations 2 года назад
Hey Jennifer, parts are located here: idealcalibrations.com/shop-all/?_bc_fsnf=1&Model=GasAlertMax+XT+II
@Grand_History
@Grand_History 5 лет назад
Does this measure O2 above 30%
@james2039
@james2039 5 лет назад
No, and any measurement above 25% is really going to be questionable with these monitors so far as I've seen.
@H6_8
@H6_8 3 года назад
witch gas the device catch?
@Idealcalibrations
@Idealcalibrations 3 года назад
Hey Hamad, this device detects Hydrogen Sulfide (H2S), Carbon Monoxide (CO), Explosives such as Methane (CH4), and Oxygen (O2).
@H6_8
@H6_8 3 года назад
@@Idealcalibrations thank you sir .. what about drager 7000 ?
@Idealcalibrations
@Idealcalibrations 3 года назад
@@H6_8 The 7000 can have any number of sensor configurations, so I wouldn't know unless I had the unit in front of me, sorry
@H6_8
@H6_8 3 года назад
@@Idealcalibrations thank you for your replay
@redstone1128
@redstone1128 3 года назад
where I can by this?
@Idealcalibrations
@Idealcalibrations 3 года назад
Hey there. I don't have it for sale on our web store, but if you want to give me a call at 734-234-5042 I'd be happy to sell you one. Or email the request to support@idealcalibrations.com. Believe we have 1 or 2 currently in stock
@user-ii4zx7le7u
@user-ii4zx7le7u 8 лет назад
здраствите у миня ест такой апарат 6 чтук
@Dianapadron81
@Dianapadron81 2 года назад
You have a very pleasant voice. You should make more instructional videos.
@Idealcalibrations
@Idealcalibrations 2 года назад
Ha! I appreciate the compliment. Voice on here was from a much younger man 😂🙏 We've done a few more since then. What sort of topics would you like to see more videos about? Probably going to do a new content run in a couple of weeks once the new website is finished.
@Dianapadron81
@Dianapadron81 2 года назад
@@Idealcalibrations well good voice! 😂 There are just some voices that don’t capture my attention. I can’t learn from those. 🤷🏻‍♀️
Далее
Altair 4XR Calibration Guide by Ideal Calibrations
8:56
QRAE 3 Calibration Guide by Ideal Calibrations
5:33
Просмотров 10 тыс.
Fixed gas Detector -Riken Keiki 2002K basic operation
7:11
Calibration manual GMI PS200
4:42
Просмотров 10 тыс.
How to Bump Test a BW Max XTII
2:01
Просмотров 4,5 тыс.