اهلا بحضرتك يا دكتوره، start by oxygen weaning, till reaching 25% or less then wean pressure till reaching 5 Cm H2O at this time and the baby stable wean either to room air or to nasal cannula
PEEP is positive End Expiratory Pressure term used when you have ventilator rate and PIP peak inspiratory pressure. CPAP continuous positive Airway Pressure, مستوى CPAP فى الجهاز ده يحدده مدى عمق Expiratory limb تحت الماء
Very good important question : There are many studies compared the efficacy of different types of bubble CPAP, including F&P and low resources bubble CPAP, F&P is better in pressure delivery and less work of breathing F&p Interface(the prong) has one of the lowest resistance and better pressure delivery. The low resource CPAP is very economic and cheap in comparison to F&P WHICH IS SUITABLE FOR OUR RESOURCES, Although high resistance especially with nasopharngeal tube and the variability of the pressure delivered, It is working.
اهلا د. اسماء، This depends on how much Fio² you need to start with, 4 liter O2+4 liter air give almost 60% oxygen,. As we usually start with 40% oxygen so start with 2 liters oxygen and 6 l air. Remember that air has 21% oxygen