Тёмный

HillRom Versacare P3200 W/ Bed Exit Wrench Illuminated 

Bearded Biomed
Подписаться 896
Просмотров 2,6 тыс.
50% 1

Опубликовано:

 

4 окт 2024

Поделиться:

Ссылка:

Скачать:

Готовим ссылку...

Добавить в:

Мой плейлист
Посмотреть позже
Комментарии : 5   
@TherealAlisaApplewhite
@TherealAlisaApplewhite 7 месяцев назад
This was the best video ever
@beardedbiomed
@beardedbiomed 7 месяцев назад
Thank you
@TherealAlisaApplewhite
@TherealAlisaApplewhite 6 месяцев назад
Your Welcome@@beardedbiomed
@j.cobainlennon9671
@j.cobainlennon9671 2 месяца назад
What if the lights don’t flash but the head still comes down?
@beardedbiomed
@beardedbiomed 2 месяца назад
If the head motor moves after jumping those pins it's typically the motor but if you jump the pins nothing moves and lights all illuminate I've found it to be a logic board issue. Really depends on revision as well as some revisions have two boards while this version I was looking at in the video was the singular board that controls everything at the end of the day running through this with Hillrom tech support is always your best call