The rotation resistance of a handpiece depends on the handpiece model, condition, and internal wear of the handpiece gears. This Control Unit incorporates an automatic function to recognize the level of the resistance of the attached handpiece and motor.
1.Attach Gear Ratio 20:1 handpiece to the motor and make sure the display is in the 20:1 Gear mode.
2.Attach the Calibration Bur to the handpiece.
3.Press the [CAL] key for about 2 seconds. There will be a beep and the display will switch to Calibration Mode.
4.Hold the motor in your hand and make sure that "L" is displayed on the torque display. Press the [CAL] key while allowing the bur to spin freely. This calibration step may take approximately 10 seconds and is complete when there is a beep and "donE" appears on the display.
5.Make sure "H" is displayed on the torque display, plug the calibration bur into the Torque Load at the top of the control unit. Push [CAL] key (This calibration step is done while applying torque load, so securely grasp the handpiece.) The calibration bur will begin to rotate. This Calibration step is complete when there is a beep and "doneE" appears on the display.
6.Remove the bur from the Torque Load at the top of the unit and then press the [CAL] key (Calibration bur should not touch anywhere). The bur will begin to rotate at a slow speed. After rotating slowly for 8 seconds, the bur will rotate at high speed for 8 seconds. There will a beep when calibration is complete. Calibration mode will finish automatically.
Please reference the SurgicPro Operation Manual prior to use.
Contact Brasseler USA for more details at 800-841-4522 or visit online at www.brasselerusa.com

Наука

Опубликовано:

 

25 май 2014

Поделиться:

Ссылка:

Скачать:

Готовим ссылку...

Добавить в:

Мой плейлист
Посмотреть позже
Комментарии : 1   
@pablovo775
@pablovo775 Год назад
Fiz exatamente igual. Tudo conectado certinho. As vezes vai as vezes n vai. Deve ser mal contato da peça. Coisa nova dando problema.
Далее
NSK Surgic Pro Implant Motor Set Up and Functions
9:17
Step-by-Step - Montagem do Motor Neodent
6:35
Просмотров 13 тыс.
How To Set Up A W&H Implantmed
3:53
Просмотров 24 тыс.
CALIBRACIÓN SURGIC AP Y SURGIC PRO
4:12
Просмотров 4,6 тыс.
Programação do motor NSK para implantes
4:18
Просмотров 6 тыс.
Nobel Biocare Torque Wrench
7:01
Просмотров 124 тыс.