Тёмный

Vacuum Pump Gas Ballast Operation 

Pope Scientific, Inc.
Подписаться 363
Просмотров 39 тыс.
50% 1

This video explains the use of a gas ballast on a rotary vane vacuum pump to remove volatiles from the oil used for sealing the pump.

Наука

Опубликовано:

 

12 авг 2020

Поделиться:

Ссылка:

Скачать:

Готовим ссылку...

Добавить в:

Мой плейлист
Посмотреть позже
Комментарии : 13   
@GnosisMan50
@GnosisMan50 Год назад
thank you for sharing your video. I have a question: When you have the ballast set to zero and you let it run to remove volatiles, how long should you let it run at zero and what is the difference when running the pump at 1 or 2?
@PopeScientificInc
@PopeScientificInc Год назад
When the ballast is set at zero there is no air introduced, a small amount is introduced when set to one and more when set to two. We recommend setting the ballast to two for at least 30 minutes after processing, but a longer amount of time is preferable. When the pump is set to zero, it will not be optimized for removing volatiles, but instead will pull the deepest possible vacuum. If very strong vacuum is not required, the gas ballast can be left open (set to one or two) while processing.
@disgraziato5473
@disgraziato5473 2 года назад
Hello - good video. I have an Edwards D40B vacuum pump. We exhaust our ballast directly to a fume hood to reduce the odour. The distance is long, and I think residual off gas is coming back into the oil and decreasing its life. Do you have any suggestions as to how I can prevent odour and at the same time increase the longevity of our oil? I was thinking of using a trap of some sort?
@PopeScientificInc
@PopeScientificInc 2 года назад
Hi, Thank you for the compliment! The first thing to check is whether or not you have any suction (even slight) at the outlet of the vacuum pump / inlet of your line that goes to the fume hood. If you create back pressure at the pump outlet, this can reduce the ultimate vacuum achieved by the pump. If this doesn’t clear things up, feel free to give us a call and we will try and provide some more clarity!
@RajasekarRamu-pq7bf
@RajasekarRamu-pq7bf 10 месяцев назад
Smoke coming continued any issue..pls suggestions?
@albertrios5168
@albertrios5168 7 месяцев назад
We are using a Pfeiffer Duo 3 for a Scanning Electron Microscope. we only look at solid items. Should I keep the ballast open at 1 or closed? (ours shows 0 or 1). Or should I have it open until a specific vacuum range is achieved and then close it? Thank you.
@PopeScientificInc
@PopeScientificInc 7 месяцев назад
Hello - For most rotary vane type vacuum pumps it is best to open the gas ballast prior to shutting down the pump to ensure any volatiles are removed from the pump, if this is done and the pump is being used on dry materials, you could likely start the pump with the gas ballast closed. However, it would be best to reach out to Pfieffer directly to get their expert opinion on use of your pump.
@zachreyhelmberger894
@zachreyhelmberger894 2 года назад
Oh wonderful! So I do not have to have a source of compressed gas to inject into the gas ballast port?! I have a HarvestRight freeze dryer and on the side of the pump, there is a knurled knob that says "gas ballast ON OFF" in the documentation, it is called part number 46 "knob". So if I loosen that knob, it looks like it can come completely off, which I assume is not a good thing. So from finger tight, I should rotate it counter-clockwise 360 degrees or so, right?
@PopeScientificInc
@PopeScientificInc 2 года назад
It may depend on your model of vacuum pump. We utilize Edwards E2M18’s that have a knob for the gas ballast. This takes approximately 6 full turns before it is fully open, at which point you start to feel some resistance to opening. If you are using a different pump, a good sign that the ballast is open is that the ultimate vacuum in your system has decreased. We would recommend reaching out directly to the vacuum pump manufacturer if further clarification is required.
@jaydouglas5847
@jaydouglas5847 2 года назад
Hi Zach, What did you learn about using the gas ballast knob . Just got my harvestright unit tonight and the standard oil based pump manual says nothing about this valve, its not even shown in the booklet. Any advice/help would be greatly appreciated.
@arthamnews6779
@arthamnews6779 2 года назад
200lpm???
@arthamnews6779
@arthamnews6779 2 года назад
Lpm???
Далее
Highlights: Spanje - Gjeorgji
06:17
Просмотров 552 тыс.
How Does A Vacuum Pump Gas Ballast Work - HVAC Tip
3:45
JB Industries - Platinum Vacuum Pump - How To
6:37
Просмотров 160 тыс.
HVAC 143 Vacuum P10 Gast Ballast
4:39
Просмотров 2,1 тыс.
How to exchange the oil of the vacuum pump
4:26
Просмотров 9 тыс.
Turbomolecular Madness - A Nerve-Wracking Experience
15:26
How to check RV12 pump
3:20
Просмотров 3,4 тыс.
Edwards vacuum pump E2M40 Overhaul 01
12:47
Просмотров 4,7 тыс.
Will the battery emit smoke if it rotates rapidly?
0:11