Kangaroo Joey Pump Instructions 

Shield HealthCare
Подписаться 21 тыс.
Просмотров 597 тыс.
50% 1

Instructions for the Kangaroo Joey Pump from Shield HealthCare. Find more information about feeding tubes: www.shieldhealt...

Опубликовано:

 

15 окт 2013

Поделиться:

Ссылка:

Скачать:

Готовим ссылку...

Добавить в:

Мой плейлист
Посмотреть позже
Комментарии : 77   
@Classyo1nelife
@Classyo1nelife Год назад
Great info! My loved one was given an unsafe discharge, with no feeding tube equipment nor training for us caregivers. Smh
@ranibanister6435
@ranibanister6435 7 лет назад
thank you for showing this.... my father in law was sent home with all this equipment nobody showed us what to d
@peppapigpetra
@peppapigpetra 4 года назад
FSFRDGT YOU HJUKIJVLOMKN BBVVBKJ KVV WXDXGG HTC HHJHJVJBVBVI
@marylougilbert9040
@marylougilbert9040 8 лет назад
Mary Lou Gilbert RN Hospice : I wish I had seen this before making a home visit to observe a Live=in caregiver set up the pump. It would have made my job must easy and less frustrating. As an instructor and teacher, this video is excellent thank you
@tmhwriter
@tmhwriter 8 лет назад
Thank you for an excellent video. Shield is the best!
@trich5621
@trich5621 7 лет назад
Wrong!
@coinsbooksbyhelen9088
@coinsbooksbyhelen9088 4 года назад
Thanks so much for a great guide.
@jessicawood2972
@jessicawood2972 5 лет назад
Why does air keep getting in the line?
@rachelisaac5
@rachelisaac5 4 месяца назад
Great video
@rahiemthomas5617
@rahiemthomas5617 3 года назад
Feeding pump is saying "feeding is complete", but there is still plenty of formula left in the bag.
@thaliaoneal3847
@thaliaoneal3847 5 лет назад
I have a feedingtube too I had mine since I was a baby 3 or 4 days after I was born I started doing videos about my feedingtube too
@audacitylove
@audacitylove 3 года назад
Thank you this video saved the day!
@barbaramarshall2536
@barbaramarshall2536 4 года назад
Do you make the the little stands for the milk to hang in? My son lives at home with us and we don't have a pole at home. Thankyou
@jaynelockhart1825
@jaynelockhart1825 3 года назад
So if the feed rate is 15ml/20mins you set the rate to 45ml/hr with a feed volume of 15ml??
@lauradecker473
@lauradecker473 7 лет назад
I tried to fill the bag with just the formula (has the flush bag) and it wouldn't keep priming. Do you always have to have the water in the flush bag for it to keep priming?
@Shieldhealthcare
@Shieldhealthcare 7 лет назад
Hi Laura, I believe that if you just want to prime the feed bag vs the flush bag you may need to you use a different command. Instructions in the manual specify: "If a Feed & Flush Set is used, remember to first press HOLD TO PRIME FLUSH until the flushing fluid has reached the valve, and then press HOLD TO PRIME FEED until the feeding fluid has reached past the valve and down to the distal end connector." Try using HOLD TO PRIME FEED to see if that works. Hope that helps!
@collinreynolds7129
@collinreynolds7129 5 лет назад
I don’t like kangaroo joey feeding bag pump I use peptide and. Water
@pennybayne3417
@pennybayne3417 7 лет назад
Thank You! This was great!! I am wondering if the pump has a feature to stop if the tube becomes disconnected. This is happening often on an overnight feeding and the gentleman gets saturated.
@Shieldhealthcare
@Shieldhealthcare 7 лет назад
Great question! The pump will alarm/stop if there is air in the line, a kink in the tubing or if the bag is empty. It won't know if the tip of the feeding set has disconnected from the patient's tube. The best way to prevent this would be to secure the tip of the feeding set into the tube's port. AMT makes a product specifically for this: AMT Clamp™ Feeding Tube Connector (www.appliedmedical.net/product/accessories/) or you can devise something similar using tape. Once your patient's doctor replaces his feeding tube with an ENFit feeding tube, the connection between the feeding tube and the feeding set will twist together, preventing disconnection. Here's an article and video on this: www.shieldhealthcare.com/community/nutrition/2016/09/12/enfit-feeding-sets-tubes-and-syringes-in-2017/
@sketchyskies8531
@sketchyskies8531 4 года назад
I was on one of these from birth until I was 9
@fannymackk
@fannymackk 4 года назад
How to take it off EZ mode it won’t let me do anything besides set feed rate
@karengrieco6825
@karengrieco6825 7 лет назад
I am trying to find out if mini backpack holds two bags, one for formula and one for flush water.
@Shieldhealthcare
@Shieldhealthcare 7 лет назад
Great question! The backpack we provide should hold both but If you're referring to the Kangaroo Mini Backpack, I'm not sure. The manufacturer should be able to answer this. Here's the number for Medtronic/Covidien: 1-800-962-9888.
@norarobie6021
@norarobie6021 4 года назад
How to change the language in the pump
@neliudisnavarro3616
@neliudisnavarro3616 10 лет назад
this video joust save me today,thanks
@Shieldhealthcare
@Shieldhealthcare 9 лет назад
Neliudis Navarro Glad to hear! Please let us know if you have any questions.
@amparogonzalez9066
@amparogonzalez9066 3 года назад
what can l do when the alarm sound every 3 minutes
@etsout
@etsout 7 лет назад
Can you use regular tap water for flushes or do you have to use sterile water?
@Shieldhealthcare
@Shieldhealthcare 7 лет назад
Hi etsout. Thank you for your question. We have an article about exactly your question here: www.shieldhealthcare.com/community/nutrition/2013/09/27/ask-the-rd-what-kind-of-water-should-i-put-down-my-feeding-tube/ Hope that helps, and feel free to ask more questions!
@rayvasquez1977
@rayvasquez1977 7 лет назад
I'm getting a rotor error while priming manually or auto priming. it will start to prime for about 3 seconds than error occurs. I release the button than repeat the process until it's primed. any idea on why I get this error while priming? I've changed out the bags, powered it on/Off with no resolution. Is there a way to factory reset to default in hopes of getting rid of error?
@Shieldhealthcare
@Shieldhealthcare 7 лет назад
Hi Ray, thanks for your question. According to the operators manual, the ROTOR ERROR screen appears during RUNNING or PRIMING, when the pump has detected an unusual operating condition for the rotor. It generally results from a problem with the pump set tubing around the rotor. Check that the pump set is not damaged, and re-load the tubing around the rotor. Press CONTINUE to restart.If the error cannot be resolved, press POWER to stop operation of the pump and put a different pump into service. In other words, if reloading a new set/bag (and maybe turning the pump off/on and unplugging/plugging the power cord) doesn't work, you may need to contact your medical supply company for a replacement pump or the manufacturer for replacement. Here's the number for Medtronic/Covidien: 1-800-962-9888. Hope that helps!
@jessicabenning5808
@jessicabenning5808 6 лет назад
How the hell do ypu turn the damn thing off?? I power down it counts down like its gonna shut off the turns bck on. Wth
@amycarrera8969
@amycarrera8969 6 лет назад
Thanks for your question. The operators manual states to press and hold the POWER button until the display turns off. If you are still having trouble there may be an issue with the pump. Try contacting your medical supply company or the pump manufacturer's customer support # 877-227-3462 Option 1. Hope that helps!
@elizabethwyneken4739
@elizabethwyneken4739 4 года назад
You’re holding the power button to long
@manghugo65
@manghugo65 5 лет назад
How to go back to English language ? This one mine went into Czech mode ?
@Shieldhealthcare
@Shieldhealthcare 5 лет назад
1. Restart pump 2. Press second (from the top) button down 3. Press fifth (bottom) button down 4. Press third button down 5. Use first and second buttons to choose language 6. Press fifth button twice
@dorothyottinger6214
@dorothyottinger6214 8 лет назад
My pump needs calibration ..is there any way to calibrate them
@jenortega4427
@jenortega4427 9 лет назад
Hi I have a question,.this sounds crazy but..anyhow ''Do you have to hook the G tube to the patients then start setting the pump? pls enlighten me
@Shieldhealthcare
@Shieldhealthcare 9 лет назад
+Jen Ortega Hi Jen. Thanks for commenting. To answer your question, you just need to have the patient connected to the feeding pump before your hit start on the pump. Please let us know if you have other questions. Thanks!
@agjc1221
@agjc1221 8 лет назад
+Jen Ortega You can set everything up without it being attached to the patient's G tube. You definitely don't want to prime the line while it is attached, because all you will do is pump air into the patient's stomach. My advice is to configure the settings (dose & rate), then prime, then attach to the patient to run it.
@ianjuarez4579
@ianjuarez4579 7 лет назад
Jen Ortega if you have the tube attached to the patient during priming, it will force air into their tube, which can be dangerous and cause discomfort.
@lanthy113
@lanthy113 6 лет назад
Thanks. Great video
@stephanieadlerm
@stephanieadlerm 8 лет назад
What is the part number for the a/c adapter? Mine didn't come with one and the manual doesn't have this information. I need to buy one so I can start using it. Is it just a regular a/c adapter? Thanks
@amylongcarrera1402
@amylongcarrera1402 8 лет назад
+stephanieadlerm Thank you for your question. I don't believe there is a separate part# for the adaptor as it should come with the pump, which cannot be used without it. Please contact the company that provided the pump to you. They should send you an adaptor. Thanks!
@teressapoe9694
@teressapoe9694 7 лет назад
how do I get it to run more then an hr
@Shieldhealthcare
@Shieldhealthcare 7 лет назад
Hi Teressa, If you mean you want to calculate the rate to run formula over a period that is longer than an hour, take the total volume you want delivered and divide it by the # of hours you want it to run. For example, if you want to run 250ml over 2 hours, then 250ml divided by 2hrs = 125ml/hr and this is the rate you will set on the pump.
@D4NkQu33n420
@D4NkQu33n420 8 лет назад
pump is showing an error while continuously feeding. the ml fed stops counting until you press hold then stops again. is there anyway to reset the pump?
@amylongcarrera1402
@amylongcarrera1402 8 лет назад
+Phoenix bennu Hi there, here goes! If you are experiencing a FLOW ERROR on the screen, the manufacturer’s instructions are as follows:The FLOW ERROR screen appears when either the formula or water is no longer being delivered due to a clog, air bubble or kink in the tubing between the pump and the patient. Check the tubing to find the reason for the blockage. If nothing is visible and there is no clog in the patient’s feeding tube, check the valve pocket in the pump loading set area for dampness or dust, which may be a cause for false error. Clean and dry this area and start the feed again. If still not resolved, try loading another set or feeding bag, prime it and press continue to restart the feeding. You may also have a dose set so that the pump is programmed to stop running when a certain amount of formula or water has been delivered or you may have the pump set on intermittent feedings. Check your settings. You may want to have the pump in EZ Mode. In this very basic mode the pump starts when you start the feeding and doesn’t stop until you stop it or it runs out of formula or water. You don’t set a volume to be delivered, or dose. You know your pump is in EZ mode if there is an EZ next to the battery symbol in the lower right hand corner of the screen while the pump is running. To change or check the mode, access the Biotech Options screen in this way: Turn the pump off. Now turn it back on and while the kangaroos are jumping across the screen, press and hold the top arrow button. Press the arrow for Biotech Options and check or un-check EZ Pump Mode depending on which you prefer. Now press done. If you want to be able to set the dose , or volume to be delivered, you cannot be in EZ mode. To set the dose, press More on the opening menu to access the More Options menu and choose continuous, then done. You can adjust the feed rate and dose (volume to be delivered, or VTBD) in the Adjust Feed screen. I hope that helps! If you need more information, you can access a PDF of the operator’s manual here: www.covidien.com/imageServer.aspx?contentID=35727&contenttype=application/pdf
@D4NkQu33n420
@D4NkQu33n420 8 лет назад
Amy Long Carrera thank you so much. I just kept hitting continue and it eventually worked. I will save this link.
@galinastruk1485
@galinastruk1485 Год назад
In.vendita macchina peg
@juanberrios3097
@juanberrios3097 4 года назад
How can I turn off the “feeding complete” notification. It wakes us up in the middle of the night I’d appreciate help with this. Thank you
@Tatiana-gc9zh
@Tatiana-gc9zh 4 года назад
You have to turn off the pump
@kauigirl808
@kauigirl808 4 года назад
@@Tatiana-gc9zh But it keeps doing that ALL the time.
@WyattTehMyLittlePonyFan86
@WyattTehMyLittlePonyFan86 7 лет назад
I wish the name of the pump was Vauxhall
@brandiwilson7493
@brandiwilson7493 6 лет назад
Why clog in line Downstream of pump error keep coming on
@amycarrera8969
@amycarrera8969 6 лет назад
A Flow Error, or downstream occlusion between the pump and the patient could be related to a kink or air in the pump bag tubing. If clearing the line or replacing the pump bag doesn't solve the problem, then check for a clog in the patient's feeding tube or extension set. Here's some info on resolving a clogged feeding tube: www.shieldhealthcare.com/community/nutrition/2015/12/03/tube-feeding-how-to-unclog-your-feeding-tube-video/. Hope that helps!
@LastStandin2011
@LastStandin2011 9 лет назад
How do you set the rate? For instance if i do not want the feed to spread over an hour, how can i make it go for like 15 minutes or 30 minutes. Because my son has a G tube and he also feeds by mouth as well. Some times he does not take the whole feed. and the rest is given through the pump. he takes 120 ml, so if he take half. do i really have to give the other 60 over and hours time?
@Shieldhealthcare
@Shieldhealthcare 9 лет назад
Wavy Henchmen Hi there! Here is the answer from our Corporate RD: If you want to cut the time in half to deliver the 60mL, double the rate. You will put 60mL in the bag and set the pump to run at 120mL/hr for it to finish in half the time (30 minutes). Here’s the math: To calculate the rate to set on the pump for anything other than one hour: -Convert hours to minutes. Example: 1 hour = 60 minutes -Divide the number of minutes you want the pump to run into 60 min. Example: 60 minutes divided by 30 min = 2 -Multiply that number by the amount of formula that you want to deliver. Example: 2 x 60mL = 120mL/hr. This is the rate you will set on the pump. If you put 60mL in the bag and run it at 120mL/hour it will finish in about 30 minutes. This also works if you want to deliver 60mL over more than one hour, for example for 90 minutes: 60 min divided by 90 minutes = 0.67 0.67 x 60mL = 40mL/hr. This is the rate you will set the pump to run for 60mL to finish in about 90 minutes. Please let us know if you have any questions! Thanks.
@chantellcoleman1753
@chantellcoleman1753 7 лет назад
did is a good one
@WyattTehMyLittlePonyFan86
@WyattTehMyLittlePonyFan86 7 лет назад
I have a g tube
@ventilator98
@ventilator98 10 лет назад
How do I get into the system settings menu?
@Shieldhealthcare
@Shieldhealthcare 9 лет назад
Peter Foster Hi there! To answer your question, or if anyone else has this question: If the pump is in EZ Mode, this mode does not allow you to set a volume or intermittent feeds. All you can do is set the rate. If you would like to do more than set the rate, you will need to take the pump out of EZ Mode. Here are instructions on how turn EZ Mode on and off, which are on page 20 in the pump manual (www.covidien.com/imageServer.aspx?contentID=35727&contenttype=application/pdf) 1. Turn the pump on and while the kangaroos are hopping across the screen, press the top left-hand arrow. 2. Choose BioTech Screen. 3. Click on the arrow next to EZ Mode to check or uncheck EZ Mode (a check means it is in EZ Mode). Please let us know if you have any other questions. Thanks for asking!
@ventilator98
@ventilator98 10 лет назад
Can you adjust brightness?
@williammontero3048
@williammontero3048 10 лет назад
I know how to setup a joey pump, set dosing, feed rate
@williammontero3048
@williammontero3048 10 лет назад
the brightness should be in the system settings menu,, where the alarms and clock settings
@LastStandin2011
@LastStandin2011 9 лет назад
***** How do you set the rate? For instance if i do not want the feed to spread over an hour, how can i make it go for like 15 minutes or 30 minutes. Because my son has a G tube and he also feeds by mouth as well. Some times he does not take the whole feed. and the rest is given through the pump. he takes 120 ml, so if he take half. do i really have to give the other 60 over and hours time?
@williammontero3048
@williammontero3048 9 лет назад
set the dose for 120 ml, then set the feed rate at 480 ml per hour. it should stop at 120 approx 15 minnuits. to do a 30 minnuit delivery, set the feed rate at 240 ml per hour, set dose at 120 ml, it should stop in approx 1/2 hour
@LastStandin2011
@LastStandin2011 9 лет назад
***** Thank you, i get it now. Your example really helped me out a lot. Thanks again
@edmanecastor8668
@edmanecastor8668 2 года назад
Great
@galinastruk1485
@galinastruk1485 Год назад
In vendita macchina peg
@WyattTehMyLittlePonyFan86
@WyattTehMyLittlePonyFan86 7 лет назад
6am I was at a hospital with the same feeding pump and a TV and the kids are saying about the new movie and I wanted to watch the movie and I saw niki 2015 Chevrolet Tahoe ltz with on star and
@stevenandsteve777prosatmin5
@stevenandsteve777prosatmin5 7 лет назад
Uhhh why u call it kangaroo joey
@mariamnmb1502
@mariamnmb1502 6 лет назад
If battery is low and I plug it and using it at the same time how do I stop the battery warning alarm from ringing all the time.
@mararomagnoli65
@mararomagnoli65 3 года назад
In italiano
@barbarabarnaby3560
@barbarabarnaby3560 7 лет назад
Ethans pump
@bernardboyle3617
@bernardboyle3617 4 года назад
Can g
Далее
Joey Feed and Flush
22:44
Просмотров 38 тыс.
Tube Feeding Using a Pump
7:14
Просмотров 150 тыс.
Classic Italian Pasta Dog
00:20
Просмотров 2,1 млн
女孩妒忌小丑女? #小丑#shorts
00:34
Просмотров 10 млн
PROGRAMMING A KANGAROO PUMP
6:23
Просмотров 29 тыс.
Kangaroo Joey Dual Bag Set Up
7:11
Просмотров 129 тыс.
Kangaroo ePump In Service
14:58
Просмотров 337 тыс.
Infinity Pump Instructional Video
20:17
Просмотров 134 тыс.
Kangaroo OMNI™ Troubleshooting Video
13:30
Просмотров 2,6 тыс.
Kangaroo OMNI™ Inservice Video
15:31
Просмотров 11 тыс.
EnteraLite Infinity Pump Training
5:44
Просмотров 47 тыс.
Kangaroo Tube Feeding Pump
11:44
Просмотров 52 тыс.
Enteral Feeding Pump 1
10:57
Просмотров 119 тыс.