Тёмный

AMSCO™ 2532 Washer/Disinfector In-Service | STERIS 

STERIS Healthcare United States
Подписаться 3,3 тыс.
Просмотров 325
50% 1

Visit the STERIS University course to earn a Certificate of Completion or CE Credit here: university.steris.com/course/...
Cette vidéo propose des sous-titres français, en sélectionnant dans le menu Paramètres RU-vid.
This video shows the proper operation of the AMSCO™ 2532 Washer, which is part of the 2500 series washers, and includes:
• 0:00 Introduction
• 1:38 Cleaning & Disinfection Practices
• 2:37 Washer Components
• 6:57 How to Run a Cycle
• 10:48 Unloading the Washer
• 13:42 Routine Care & Maintenance
Refer to the equipment operator manual for complete information.
To learn more about the AMSCO™ 2532 Washer/Disinfector and request a quote, please visit: bit.ly/3VBqDhZ
Some of the key information included in the in-service video includes:
Components and controls:
The unit is equipped with one or two front-loading drop-down chamber doors. The doors are provided to facilitate loading and unloading of the machine. There is an icon-based touchscreen for performing various cycle operations on the load side. On the unload side are indicator lights, two acknowledgment push buttons, an emergency stop, and, if selected, an optional printer. If you're using the single-door model, your unit has all load-side features.
Running a Cycle:
Pre-programmed cycles and cycle icons are used to start a cycle and track its progress.
To initiate a cycle, select “Cycle” from Ready mode to see pre-programmed options. Once the cycle is finished, a signal sounds and the control screen displays the “Cycle Completed” screen. A detailed printout is generated documenting the cycle parameters.
Routine Care & Maintenance
To keep the washer in excellent condition, it’s important to clean and maintain it regularly. A maintenance schedule can be found in the Operator Manual. Some maintenance activities should be performed by a qualified service technician.
Certain preventive maintenance activities must be done by a STERIS service technician or a STERIS-trained service technician in your facility. These activities include HEPA filter verification and replacement, Squeeze Tube verification, Chemical Injection Pump verification, and inspecting and cleaning the Rotary Spray Arm assemblies.
Professional standard organizations recognize the importance of monitoring your mechanical and chemical cleaning process. STERIS recommends using the Verify® All Clean Test Washer Indicator on a daily basis to monitor washer cycles.
After the last cycle has been completed each day, you’ll need to clean & rinse off the Debris Screen.
Weekly Maintenance:
• Clean the outside of the washer
• Inspect and clean rack spray arms and tips
• Run the decontamination cycle
Some maintenance activities occur on an as-needed basis. This includes replenishing the processing chemicals and replacing the printer paper and ink cartridge.
In addition to routine maintenance, regularly scheduled preventive maintenance is essential for ensuring peak performance and avoiding unscheduled downtime. To schedule a maintenance appointment with a STERIS service technician, call 800.333.8828

Опубликовано:

 

23 июл 2024

Поделиться:

Ссылка:

Скачать:

Готовим ссылку...

Добавить в:

Мой плейлист
Посмотреть позже
Комментарии    
Далее
80 Year Olds Share Advice for Younger Self
12:22
Просмотров 1,3 млн
Strong cat !! 😱😱
00:19
Просмотров 2,7 млн
Decontamination and cleaning
9:40
Просмотров 149 тыс.
The Problem With These Headlights
17:55
Просмотров 849 тыс.
Doff Steam Cleaning System Explained
9:53
Просмотров 5 тыс.
How Do You Steer a Drill Below The Earth?
14:53
Просмотров 4,1 млн
Steris Ultrasonic Inservice
9:00
Просмотров 753