Basic Procedure to Setup the SEM after Loading Sample 

The Kavli Nanoscience Institute at Caltech
Подписаться 4,3 тыс.
Просмотров 4,9 тыс.
50% 1

After the chamber has reached suitable vacuum, follow these steps to set the voltage and current, start the beam, find your site of interest, and get the sample to your desired working distance. Alignments are then covered in subsequent videos (start alignments here: • Basic SEM Alignment (S... )
** Please note the setup in this video was performed on flat samples. Non-flat samples have limitations on setup (instrument damage can result otherwise in some circumstances) which is not discussed in this video **
When moving across the sample, consider if adjusting the stage height is required: samples are seldomly mounted flat or are truly even. It is possible to end up in a different working distance position (or to fall out of eucentric height if that was set) by moving along the sample.
___
Disclaimer: This video is intended to serve as supplemental information and cannot replace in-person instrument training. Microscope operators are encouraged to consider which parameter setting is most suitable for their measurements. Caltech and KNI are not responsible for how the presented information is translated for use in other facilities.
For information or questions about this video, contact kni@caltech.edu

Наука

Опубликовано:

 

12 окт 2017

Поделиться:

Ссылка:

Скачать:

Готовим ссылку...

Добавить в:

Мой плейлист
Посмотреть позже
Комментарии    
Далее
Scanning Electron Microscope (SEM)
13:27
Просмотров 84 тыс.
SEM: Zeiss Sigma HD Scanning Electron Microscope
9:22
Cosmology in Crisis? Confronting the Hubble Tension
36:26
The Man Who Solved the World’s Hardest Math Problem
11:14
Optimizing SEM Imaging
25:16
Просмотров 7 тыс.
Robotaxis | Big Ideas 2024
9:41
Просмотров 70 тыс.
Flow Cytometry Analysis
13:24
Просмотров 97 тыс.
Я УКРАЛ ТЕЛЕФОН В МИЛАНЕ
9:18
Просмотров 97 тыс.