How to Clean an HPLC Column in Daily & Deeper Cleaning 

Filtrous
Подписаться 880
Просмотров 15 тыс.
50% 1

In this video, we are going to teach you how to clean your column. 🧼
This is especially important to do if you add buffers or additives to your mobile phase. For instance, most people use ammonium acetate, formic acid, etc. Over time, if the column isn’t flushed out regularly, precipitates can form and can plug your column up.
We are going to show you 2 techniques to extend your column life as much as possible.
The first method is your daily maintenance method. If you have a solenoid valve in your pump, you can select up to 4 solvents. A and B can be used as your mobile phase for regular operations. C and D can then be selected to be used as your solvents for cleaning. C and D would normally consist of water and some kind of organic solvent like methanol. You can simply do a gradient after your last run of 5% to 95% methanol for over an hour.
If you feel like your column needs a deeper cleaning, it would be best to do a reverse flush. The method for reverse flush would be to
95% water and 5% methanol for 10 column columns
Do a gradient of 5 to 95% methanol for 5 column volumes
Hold 5% water and 95% for 10-column volumes
You want to run this at half your normal flow rates for your method.
The way that you know when to do deeper cleaning is to monitor the pressure build-up on your column. Always note the back pressure on your method when you first start to use your column and if pressure kicks up around 15-20% it’s probably a good idea to use the reverse flush.
When you have questions or have a topic you want to be covered let me know in the comments! 😎
Need methanol for your lab? 🛒Shop at the high-quality and reliable store Filtrous: filtrous.com/
🔗Chemicals ➡️fltro.us/3Pyjv0p
✅We arrange 20 different variations of chemicals.
🚨Clearance ➡️ fltro.us/3yOId5s 🚨
🚨🚨Don’t miss out on great deals.🚨🚨
#biotech #biology #science #hplc #clanning

Хобби

Опубликовано:

 

16 дек 2021

Поделиться:

Ссылка:

Скачать:

Готовим ссылку...

Добавить в:

Мой плейлист
Посмотреть позже
Комментарии : 18   
@fakhrulnawawi9681
@fakhrulnawawi9681 7 месяцев назад
Really helpful thankss
@thehealer8330
@thehealer8330 2 года назад
I have purchase new hplc c18 colums. I want to know how to use a new colum??. Or just start using it for routine analysis. I want your say on this
@TheJoshtheboss
@TheJoshtheboss Год назад
Excellent video
@Filtrous
@Filtrous Год назад
Thank you very much!
@kristabanks4523
@kristabanks4523 2 года назад
This is a good general usage recommendation, but what about for specialty columns? For instance, I sometimes use phenyl columns or HILIC columns. About 80% of what I analyze uses either C8 or C18 Reverse-phase columns, and I would definitely employ this cleaning method, but is there a better cleaning format for the lesser used types of columns?
@arielhu4569
@arielhu4569 2 года назад
I have the exact same question too
@user-nu6dx7sf8v
@user-nu6dx7sf8v Год назад
how about the flow rate of 10% Methanol???
@ifeoma4517
@ifeoma4517 2 года назад
Is 100 present ACN not good enough for flushing columns?
@Filtrous
@Filtrous 2 года назад
not for all situations no.
@MuhammadUsmanAkram-usman101pak
@MuhammadUsmanAkram-usman101pak 2 года назад
I’m using RP18 250 cm column with 90% 1%acetic acid and 10%MeOH on gradient where B maximum is 65%. After about 80 hours of constant use, I found peaks are moving forward for about half minute, even for standards. Can you tell me the reasons and solution. Column can work between 1-9 pH, n A phase has pH around 3.5
@Filtrous
@Filtrous 2 года назад
you're probably not equilibrating your column long enough. you might want to hold it at B longer at the end of the run or even kick it up higher and then when you bring it back to the starting conditions - let that go for a bit before ending the run.
@hoa4300
@hoa4300 2 года назад
Can you write a detailed program to clean column C18 with this method (95% water and 5% methanol for 10 column columns. Do a gradient of 5 to 95% methanol for 5 column volumes. Hold 5% water and 95% for 10 column volumes?) Tks Note: I use pharmaceutical analysis with HPLC Dionex Ultimate 3000 autosampler and Water E2695 autosampler. Example: I used ACN: KH2PO4 0,01M (40:60) pharmaceutical analysis, so How I can clean column C18 with this method?
@DavidYazdi1
@DavidYazdi1 2 года назад
You need to remove your current mobile phase and replace with methanol and water. It's important to remove the salt in your column. Hopefully you have a binary or quaternary pump where you can do a gradient. You first start with a hold of 5:95 methanol to water. Then you do gradient. Finally you hold 95:5 methanol to water.
@hoa4300
@hoa4300 Год назад
@@DavidYazdi1 Tks so much
@dhananjaygangadhare8333
@dhananjaygangadhare8333 Год назад
100 %Water is not good for column
@danielwatson7386
@danielwatson7386 Год назад
Do not follow this video. Always follow the column manufacturers manual and call their technical support if you have specific application questions. Anyone who would tell you to reverse flow through a HPLC column is basically telling you to piss way $1500. Source: I'm a field service engineer for Thermo Fisher and I work in IC/LC chromatography.
@Filtrous
@Filtrous Год назад
Hey Daniel, this is for a reverse phase column and this is in the column care guide that comes with the column. Yes, you are right that this does not apply to all columns but it does apply to this column as it is reverse phase.
@abbesdimiri8839
@abbesdimiri8839 Год назад
I invite everyone to read coran the last book of god on earth and judge it
Далее
Set Up a qPCR Plate in 3 Simple Steps
1:54
Просмотров 9 тыс.
HPLC Column Washing for Reversed Phase Column
6:26
Просмотров 2,4 тыс.
💜☀️✨
00:47
Просмотров 84 тыс.
How to Change an HPLC Pump Seal | Shimadzu LC-20AD
4:15
HPLC Chromatograph
13:19
Просмотров 16 тыс.
How to clean a C18 column?
3:33
Просмотров 3,5 тыс.
HPLC column washing for reversed phase mode.
7:16
Просмотров 6 тыс.
Deep-Cleaning a Viewer's DIRTY Gaming PC! - PCDC S2:E1
20:55
How to install an HPLC column
5:52
Просмотров 12 тыс.
Preventing LC Column Clogs
1:57
Просмотров 6 тыс.
NEGA XECHKIM LAYK BOSMAYDI
0:14
Просмотров 3,2 млн