Agar aapne calibration unit laga rakhi hai to ref air dijiye otherwise koi jaroori nahi hai open karne ki . Aap sirf cal port me connect kijiye aur zero calibration kar sakte hain
Thank you for your video.! I have a question. After finishing calibration, the next step we must do for ensuring the system working accurately or not, Sir?
Very well explained, but I am getting calibration error once I end the calibration after completing span and zero. Also, what are we supposed to do with reference port while giving Calgas?
Thanks sir, Sample gas kon si hoti hai Aur reference gas kon si. Zero gas kon si hoti h Span gas kon si hoti h Aur cylinder ke connection kese hote hai sensor se?
Sir,your video is very helpful, pleas make a video for all parameter read and change and meaning full programming step by step in this zr 402g model, i hope early my request granted in your side sir.please.
Holding means output hold ho jata hai . Timing aap adjustment kar skate hain aapne according . Cross ka matlab samajh nahi aaya . Plz send me snapshot in my telegram group .
Sir , reference air Always continues flow in calibration time , in our plant 402g O2 installed in aux boiler then we are Always reference air continues in normal running plant ya normal service , we are not off reference air in calibration time.
Reference air should be continues flow . But not in my plant ,NO problem. Actually in my plant analyzers installed outside area so atmosphere air is enough and measur value is good . So didn't use in my plant .
Agar aapke paas 21% o2 standard gas cylinder available hai to best hai . Otherwise you know that atmosphere me bhi available hai 21% gas. So aap dono me koi bhi choose kar sakte hain
Span Calibration 20.9 % me karte hain . Agar standard gas cylinder available nahi hai to aap atmosphere me bhi kar sakte hain . 8% standard se accuracy nahi aayegi .
Sir ye Boiler me use hota , flue gas me oxygen concentration measure karne ke liye . Aur baanki Maine vedio me dikhaya hai . Koi aur querry hai to ask me .
I have open the cal port in zirconia Detector in atmospheric air & reference in port closed, but its oxygen reading is not increases in Zr402g display, what is the reason