In-Office Spore Testing (Biological Monitoring) Summary:
• Sterilizer monitoring requires the use of BOTH an unprocessed Control vial and a processed Test vial.
• Both the Control and Test vial culture medium start out purple in color.
• The Test vial is processed along with a load in a sterilization cycle, whereas, the Control vial is not processed.
• Before placing both vials into an appropriate pre-set temperature incubator, both vials must be activated by crushing the sides.
• Upon full incubation, the processed Test vial should remain purple (indicating no spore growth) and the unprocessed Control vial should turn yellow (showing spore growth).
Don't Risk It! Direct questions to 1-8558-STERILE (Toll-Free) 8am-5pm M-F MST.
Crosstex International, Inc. Video Disclaimer
This video is owned by the sole property of Crosstex International, Inc. ("Crosstex") and is to be used solely for internal informational purposes. Any copying, distribution, posting or editing of this video production in any format, or any unauthorized or improper use of this video, is strictly forbidden. The information presented in this video is based on current information at the time of production. Crosstex makes no warranty or guarantee concerning the accuracy or reliability of the content of this video. For technical and regulatory guidance, please refer to applicable instructions for use and product labels with respect to any product or process described in this video and current official versions of applicable state, federal or other governmental or regulatory rules, laws, standards, policies, protocols and recommended best practices.
Content is subject to change at any time.
©2014 Crosstex International, Inc. All rights reserved.

Опубликовано:

 

22 июл 2024

Поделиться:

Ссылка:

Скачать:

Готовим ссылку...

Добавить в:

Мой плейлист
Посмотреть позже
Комментарии : 4   
@dr.harishgarre7072
@dr.harishgarre7072 2 месяца назад
What if both the control and the test vial are coming out like a clear gray after incubation?
@jessen00001
@jessen00001 Год назад
Thãnkš
@arielkhaliddelafuente4673
@arielkhaliddelafuente4673 8 лет назад
how come if the test vial and control vial is same color purple after 10hrs incubatorwhat is the problem?
@Whateverhasbeenmynameforyears
@Whateverhasbeenmynameforyears 7 лет назад
3:32
Далее
3M™ Attest™ 490M Auto Reader How To video MP4
4:45
Crosstex/SPS - Cleaning, Packaging & Sterilization
15:39
Operating an Ethylene Oxide Gas Sterilizer
4:53
Просмотров 51 тыс.
How to Change Cell Culture Media
4:55
Просмотров 102 тыс.
Sterile Processing Technician--Donna Reich
12:14
Просмотров 926 тыс.